徐锦航, 李林, 董超美, 李铁军. 大气中总烃测定直接进样法的探讨和采样器的选择[J]. 北京工业大学学报, 1983, 9(3): 129-136.
    引用本文: 徐锦航, 李林, 董超美, 李铁军. 大气中总烃测定直接进样法的探讨和采样器的选择[J]. 北京工业大学学报, 1983, 9(3): 129-136.
    Xu Jin-hang, Li Lin, Dong Chao-mei, Li Tie-jun. An Approach to the Immediate Determination Method of Total Hydrocarbon in Ambient Air and the Selection of Sampling Apparatus[J]. Journal of Beijing University of Technology, 1983, 9(3): 129-136.
    Citation: Xu Jin-hang, Li Lin, Dong Chao-mei, Li Tie-jun. An Approach to the Immediate Determination Method of Total Hydrocarbon in Ambient Air and the Selection of Sampling Apparatus[J]. Journal of Beijing University of Technology, 1983, 9(3): 129-136.

    大气中总烃测定直接进样法的探讨和采样器的选择

    An Approach to the Immediate Determination Method of Total Hydrocarbon in Ambient Air and the Selection of Sampling Apparatus

    • 摘要: 本文介绍了用活性炭低温(液氮为冷源)吸附法净化氢气以代替高纯氢发生器,用催化净化法净化载气和助燃气,从而提高了SP-2305E型气相色谱仪的灵敏度。本方法最低检出限0.02ppm,相对偏差1.5%,变动系数4.9%,分析周期10秒。文中还此较了三种催化剂的净化效果和七种材质采样器的样品保存时间。

       

      Abstract: This treatise introduces a method for purifying hydrogen by activated carbon adsorption at low temperature(liquid nitrogen as cooling agent) instead of high pure hydrogen generator, and a catalytic method for pyrifying carrier gas and combustion promotion gas, thus raising the sensitivity of gas chromatography SP-2305 E. The following are reached: minimum limit of detection -0.02 ppm, relative deviation -1.5%, coefficient of variation -4.9%, analytic period -10 seconds. In addition, this treatise has also made a comparison among the purification effects of the three catalysts and the sample storage times of the seven,sampling apparatuses of diffrent materials.

       

    /